CyberRota Analysis
This is a critical severity vulnerability with a CVSS score of 9.8. Public exploit code or proof-of-concept references have been detected in its references. It may be remotely exploitable.
Public Exploit Signal
A public exploit, PoC, GitHub repository or Metasploit reference was detected for this CVE.
Note: these links are listed for security research and verification purposes only.
Original NVD Description
Spacelabs Healthcare Sentinel versions 10.5.x and higher and 11.x.x before 11.6.0 contain an unauthenticated remote code execution vulnerability through a deprecated .NET Remoting HTTP channel exposed on port 8989 that allows attackers to perform arbitrary file read and write operations by supplying valid .NET URI endpoints. Attackers can write ASPX webshells to the IIS wwwroot directory to achieve unauthenticated remote code execution on the system. Port 8989 is not exposed in a default Sentinel installation; exploitation requires that the .NET Remoting port has been explicitly made network-accessible through deliberate configuration or network policy changes.