AUGUST 4, 2026
Live Feed
Back to database
Case File

CVE-2019-6538

CRITICAL · CVSS 9.3 EPSS 0.90% Public Exploit

Source: NVD + CISA KEV + EPSS (historical backfill) · Published 2019-03-25 · Last synced 2026-08-04

CyberRota Analysis

AI analysis pending.

Public Exploit Signal

A public exploit, PoC, GitHub repository or Metasploit reference was detected for this CVE.

Detected Signals
exploit

Note: these links are listed for security research and verification purposes only.

CVE
CVE-2019-6538
Severity
CRITICAL
CVSS
9.3
EPSS
0.90%

Original NVD Description

The Conexus telemetry protocol utilized within Medtronic MyCareLink Monitor versions 24950 and 24952, CareLink Monitor version 2490C, CareLink 2090 Programmer, Amplia CRT-D, Claria CRT-D, Compia CRT-D, Concerto CRT-D, Concerto II CRT-D, Consulta CRT-D, Evera ICD, Maximo II CRT-D and ICD, Mirro ICD, Nayamed ND ICD, Primo ICD, Protecta ICD and CRT-D, Secura ICD, Virtuoso ICD, Virtuoso II ICD, Visia AF ICD, and Viva CRT-D does not implement authentication or authorization. An attacker with adjacent short-range access to an affected product, in situations where the product’s radio is turned on, can inject, replay, modify, and/or intercept data within the telemetry communication. This communication protocol provides the ability to read and write memory values to affected implanted cardiac devices; therefore, an attacker could exploit this communication protocol to change memory in the implanted cardiac device.