CyberRota Analysis
This is a medium severity vulnerability with a CVSS score of 6.5. Public exploit code or proof-of-concept references have been detected in its references. It may lead to a denial-of-service condition.
Public Exploit Signal
A public exploit, PoC, GitHub repository or Metasploit reference was detected for this CVE.
Note: these links are listed for security research and verification purposes only.
Original NVD Description
Dräger Infinity M300 patient worn monitors with software version VG2.3.1 and earlier contain a network-based denial of service vulnerability that allows network-adjacent attackers to repeatedly trigger device reboots by sending malicious requests over the Infinity Network. Attackers can exploit this vulnerability to force the device into a fail state requiring manual restart, causing loss of wireless connectivity and interruption of patient monitoring functionality.