SEPTEMBER 1, 2026
Live Feed
Back to database
Case File

CVE-2019-25719

HIGH · CVSS 8.6 EPSS 0.13%

Source: NVD + CISA KEV + EPSS (historical backfill) · Published 2026-06-02 · Last synced 2026-08-04

CyberRota Analysis

This is a high severity vulnerability with a CVSS score of 8.6. It may lead to a denial-of-service condition.

CVE
CVE-2019-25719
Severity
HIGH
CVSS
8.6
EPSS
0.13%

Original NVD Description

Dräger Infinity Acute Care System and Standalone Infinity M540 patient monitors running software versions VG4.1.1, VG4.0.3, and lower contain network message handling vulnerabilities that allow network-adjacent attackers to spoof or tamper with data and cause denial-of-service conditions. Attackers with access to an enabled Infinity network port or physical proximity to a wireless access point can modify device settings such as alarm states or alarm limits, and overwhelm the system with incoming data causing the device to reboot and lose network functionality.