CyberRota Analysis
This is a high severity vulnerability with a CVSS score of 7.4. See the original NVD description below for full technical details.
CVE
CVE-2018-18564
Severity
HIGH
CVSS
7.4
EPSS
0.58%
Original NVD Description
An issue was discovered in Roche Accu-Chek Inform II Instrument before 03.06.00 (Serial number below 14000) and 04.x before 04.03.00 (Serial Number above 14000), CoaguChek Pro II before 04.03.00, and cobas h 232 before 04.00.04 (Serial number above KQ0400000 or KS0400000). Improper access control allows attackers in the adjacent network to change the instrument configuration.