AUGUST 4, 2026
Live Feed
Back to database
Case File

CVE-2017-12725

MEDIUM · CVSS 5.6 EPSS 0.83%

Source: NVD + CISA KEV + EPSS (historical backfill) · Published 2018-02-15 · Last synced 2026-08-04

CyberRota Analysis

This is a medium severity vulnerability with a CVSS score of 5.6. See the original NVD description below for full technical details.

CVE
CVE-2017-12725
Severity
MEDIUM
CVSS
5.6
EPSS
0.83%

Original NVD Description

A Use of Hard-coded Credentials issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. The pump with default network configuration uses hard-coded credentials to automatically establish a wireless network connection. The pump will establish a wireless network connection even if the pump is Ethernet connected and active; however, if the wireless association is established and the Ethernet cable is attached, the pump does not attach the network stack to the wireless network. In this scenario, all network traffic is instead directed over the wired Ethernet connection.